Agilent E4440A w/Preselector (PSA)

 

I have found an issue wih the PSA when using the preselector in ‘filter’ mode. 

If the PS is in filter mode and a start frequency of 30MHz or stop frequency of 1GHz is commanded (even from the front panel) the result will be a ‘compressed’ 10Hz span at either 29.999MHz or 999.999MHz. 

Example:

CF: 500MHz

Span: 1MHz

Change to:

Start Freq:  30MHz

Result:

29.999MHz to 30MHz.

NOTE:  This does not happen if the preselector is in BYPASS.

This will be evident when you are using a ’scan’ type of action such as Optimize. 

Work around:

Place two GPIB actions in the flowchart before the scan action.

In the first one:  PSA:"sense:freq:star 30MHz"

In the second one:  PSA:"sense:freq:stop 1GHz"

If I put them in the same one there was not enough time to set it.

I have contacted Agilent about this and will post here when I see the fix.

 

Joe Tannehill